ISO 7206-12:2016 pdf free
ISO 7206-12:2016 pdf free.Implants for surgery一Partial and total hip joint prostheses
plane, parallel to the frontal face reference plane, located with a certain distance to the frontal face reference plane but as close as possible to the frontal face reference plane
Note 1 to entry: The sensitivity of the deformation measurement decreases with increasing distance of the measurement plane from the front face reference plane and with decreasing distance of the measurement plane to the top of the cup.
Note 2 to entry: Within the measurement plane, the measurement points for determining the inner diameter of the test specimen can be captured. Therefore, the measurement plane can be defined so that capturing the measurement points is not disturbed by any design features of the test specimen as holes or cut-outs. The measurement points can be captured at the test specimen directly; they cannot be captured at the load frame.
The test specimen is subjected to diametrically opposite two-point loading. For the determination of short-term deformation, measurements of diameter in loading direction in a defined measurement plane are carried out prior and under loading, as well as after unloading. This deformation measurement procedure is repeated two times after rotating the specimen with rotation angles of itself, of which each rotation angle measures 120° to account for influence of asymmetric design features as fins, holes, etc.
Metal-backed modular acetabular components can deform and affect the seating of the insert. The combination of metal-backing and insert can deform and affect the tribology. So such components shall be tested in two steps: first step, testing of the metal backing alone; second step, testing of the metal backing with the appropriately mounted bearing insert.
Press-fit installation of monoblock acetabular cup components can cause deformation of the articular surface which may affect tribology. Such components shall be tested in only one step.ISO 7206-12 pdf free.