ISO 18665:2015 pdf free
ISO 18665:2015 pdf free.Traditional Chinese medicine一Herbal decoction apparatus
The hermetic container test shall be in accordance with the rules given in ISO 16528-1:2007, 7.4 and 7.5 and ISO 16528-2:2007, Clause 5.
Check the pipeline installation and the easy dismounting conditions. Place the herbal decoction apparatus on level ground and connect it to one end of the pipeline of the packaging component. Place the other end of the pipeline lower than the decoction outlet.
Add a nominal capacity of water to the decoction container, then perform the valve switch test. Open the valve to let the water flow until no more water comes out. Place a bowl at the outlet and then incline the herbal decoction apparatus more than 10 degrees towards the outlet.
Measure the water volume in the bowl with a 50 ml measuring cup and compare this value to the specified value.
Measure the minimum bore diameter in the pipeline.
Add 50 % of nominal capacity of drinking water to the decoction container. With the decocting temperature in the hermetic condition set at the maximum working temperature, and the decocting time set at 40 min, start the apparatus. Conduct the loading test in accordance with 5.3.6, 5.3.7, 5.5.1 and 5.5.2.
In the loading test, when the boiling lasts for 1 min in the non-hermetic decoction apparatus,check the margin of error values between the display temperature of the herbal decoction apparatus and 100 °C.
For the hermetic herbal decoction apparatus, observe whether the heating stops when the temperature reaches the setting temperature. The setting temperature is defined as the temperature that the operator sets on the machine before decocting. When it reaches the setting temperature, the temperature will not rise anymore.
The continuity testing of the protective grounding circuit of electrical system shall be in accordance with the rules given in IEC 60204-1:2009, 18.2.
The insulation resistance testing of the electrical system shall be in accordance with the rules given in IEC 60204-1:2009, 18.3.
The voltage test of the electrical system shall be in accordance with the rules given in IEC 60204-1:2009, 18.4.
The push-buttons of the electrical system shall be in accordance with the rules given in IEC 60204-1:2009, 10.2.
The indicator lights and displays of the electrical system shall be in accordance with the rules given in IEC 60204-1:2009, 10.3.
The wiring of the electrical system shall be in accordance with the rules given in IEC 60204-1:2009, Clause 13.ISO 18665 pdf download.