Home>ISO Standards>ISO 15096:2020 pdf free

ISO 15096:2020 pdf free

ISO 15096:2020 pdf free.Jewellery and precious metals一Determination of high purity silver一Difference method using ICP-OES
The sample is weighed and dissolved in nitric acid to prepare a 10 g/l solution (higher concentration is used for fineness of and above 999,9 %o). The suspension, which can be present in that solution, is isolated by centrifugation or microfiltration and dissolved in aqua regia. Both nitric and aqua regia solutions are analysed separately by ICP-OES and the total content of each impurity (see Table A.1 for wavelengths) in the sample is obtained by adding together the results of the two analyses. The silver content is obtained by subtraction of the total content of impurities in the sample from 1 000 %o. For the determination of fineness of and above 999,9 %o, modifications described in Annex B shall be applied.
Weigh (500±10) mg of the sample portion to the nearest 0,1 mg, transfer into a 50 ml beaker, and add 7,5 ml of H20 and 7,5 ml of HNO3 (5.2). Heat gently until complete dissolution of the sample and continue to heat to expel the nitrogen oxides. Add 10 ml of water. After cooling to room temperature, transfer the solution from the 50 ml beaker to the centrifuge tube. Centrifuge the mixture at 3 000 r/min for 5 min and transfer the liquid phase into a 50 ml volumetric flask. Add 5 ml of water to the residue, centrifuge the mixture again at 3 000 r/min for 5 min more, and add the newly formed liquid phase to the 50 ml volumetric flask. This operation is repeated two extra times. Make up the volume of the volumetric flask water to 50 ml, and mix thoroughly.
Alternatively, after cooling to room temperature, filter the solution through a microfiltration system (6.5),add the filtrate to a 50 ml volumetric flask and make up the volume of the solution with water to 50 ml,and mix thoroughly. Analyse this solution without delay using the calibration solutions specified in 8.3.Add 2 ml of freshly prepared aqua regia (5.3) to the residue after centrifugation or microfiltration (if microfiltration is used, the filter shall be dissolved, too]. Heat until complete dissolution and continue to heat to expel the nitrogen oxides. Cool to room temperature, transfer to a 5 ml volumetric flask, rinse with water, and make up with water to 5 ml. Analyse this solution using the aqua regia calibration; solutions specified in 8.4.ISO 15096 pdf free download.

Related standards